One table with every technical figure HANS Korea publishes for the NiTi Brush, each with its source: the Instructions for Use (document HKIFU-001, Rev. 07, 22 May 2023), the raw-material data on this site, the general brochure, and the regulatory certificates. Figures that are not published are marked “not published; on request” rather than estimated. The IFU is the binding document for use. The NiTi Brush is also known as a titanium brush or Ti brush, was formerly named I.C.T. (Implant Cleaning Technique), and is sold in the United States by HANS Korea's distributor under the name RotoBrush-Titanium.
Specifications all published figures, with sources
| Device | |
|---|---|
| Device name | NiTi Brush for peri-implantitis (NiTi Brush™, unregistered trademark). Former model name I.C.T.; Korean brand name 싹쓸이 (Ssakssuri); United States distributor label RotoBrush-Titanium.Source: Literature page instrument lines; company history. |
| Device type | Single-use rotary dental instrument for mechanical debridement of dental implant surfaces; GMDN 57955 (dental implant debridement brush).Source: company page, regulatory status. |
| Intended use (IFU) | “mechanical debridement and cleaning of titanium dental implant surface contaminated by osseous defect of plague, calculus or any other foreign body materials resulted from peri-implantitis”; for use in the dental office by dentists, dental hygienists and trained dental assistants. Indication: peri-implantitis.Source: IFU HKIFU-001 Rev. 07, Intended Use / Indication. |
| Models | NANO (pointed, bound tip; one thread exposed) · POCKET (open head; two to three threads) · POCKET SHORT (POCKET head on a short shaft; narrow posterior sites, limited mouth opening) · OMEGA (wide, splayed head; more than three threads, deep pocket). Same bristle, same shaft diameter; the model name is marked on the shank.Source: product page; HANS Korea model criteria. |
| Manufacturer | HANS KOREA CO.,Ltd., 314Ho, 107 Garam-ro 116beon-gil, Paju-si, Gyeonggi-do 10896, Republic of Korea. Manufactured since 2010.Source: IFU; company page. |
| Materials | |
| Bristle alloy | Nickel-titanium (Ni-Ti) shape-memory alloy wire.Source: product and raw-material pages. |
| Bristle wire diameter | About 0.095 mm.Source: titanium-brush checklist page; implant-surface-brushes page. |
| Modulus of elasticity (bristle alloy) | 28–41 GPa (titanium 115 GPa; stainless steel 210 GPa, published handbook values). Material properties are not clinical outcomes.Source: raw-material page. |
| Hardness, Vickers (bristle alloy) | 303–362 (titanium 200–340; stainless steel 600–610).Source: raw-material page. |
| Shank material | Stainless steel to ISO 1797-1.Source: HANS Korea general brochure (“Material of Shank: SUS (ISO 1797-1 / metal)”). |
| Shank and head dimensions | Not published; on request. |
| Drive and use | |
| Drive | “CE marked rotating implant motor (handpiece)”; “Any handpiece that has the deceleration rate of ‘20:1’ is recommended.” Not intended for a high-speed turbine or a straight handpiece.Source: IFU, How to connect. |
| Speed | “Recommended RPM for operation is 600 RPM ~ 1,200 RPM” (speed at the brush, after the reduction ratio).Source: IFU. |
| Irrigation | “Rinse extensively with sterile saline solution throughout the whole brushing time.”Source: IFU, How to Use. |
| Brushing time | Not specified in the IFU. Published trials report about 2–3 minutes at 600 rpm or use at up to 1,200 rpm (see the evidence archive). |
| Several implants, one patient | Immerse the brush in 3% hydrogen peroxide and rinse with sterile saline before the next implant (IFU recommendation).Source: IFU. |
| Operating-room temperature | 15–30 °C.Source: IFU, Checklist before Use. |
| Supply, sterilisation, storage | |
| Supplied | Non-sterile. Packaging: single pack (one brush in a bur holder, sealed blister) and 4-piece pack.Source: IFU (“NiTi Brush is non-sterile product”); product page, How it is supplied. |
| Sterilisation before use | Autoclave · configuration: wrapped · 134 °C · 2.5 bar · exposure 15 minutes · dry time 20 minutes. “After sterilization, use the NiTi Brush immediately without re-packing or re-storage.”Source: IFU, Sterilization Method & Condition. |
| Reuse | “NiTi Brush is for single-use only. Do not re-use the NiTi Brush.” Discard after use (cross-infection risk); dispose of as a disposable medical device under local rules.Source: IFU, Warning / How to Disposal. |
| Storage | 15–30 °C; humidity 1–75%; away from sunlight; do not load other materials on the package.Source: IFU, How to Storage. |
| Shelf life | 5 years from the date of manufacture.Source: IFU, Expiry date. |
| Contraindications and side effects | See Safety, contraindications and limits (titanium or nickel hypersensitivity; others as listed in the IFU). |
| Regulatory | |
| Classification | Class IIa medical device (EU).Source: company page. |
| CE marking | EC certificate 2195-MED-1734601, issued under Council Directive 93/42/EEC (MDD) as amended by 2007/47/EC; validity extended under Regulation (EU) 2023/607 (Article 120 MDR) to 31 December 2028. Notified body: SZUTEST Konformitätsbewertungsstelle GmbH. MDR transition in progress.Source: company page, regulatory status. |
| EU authorised representative | Eunitor GmbH, Kennedydamm 5, 40476 Düsseldorf, Germany (SRN DE-AR-000005081).Source: IFU; company page. |
| Manufacturer SRN | KR-MF-000018237.Source: company page. |
| United States | FDA establishment registration and device listing. Registration and listing is not FDA approval or clearance.Source: company page; titanium-brush checklist. |
| Republic of Korea | 식약처 신고 (notification to the Ministry of Food and Drug Safety, MFDS) — number on request. In Korea, use follows the notified intended purpose and the Korean IFU.Source: HANS Korea. |
| Instructions for Use | Document HKIFU-001, Rev. 07 (22 May 2023); multilingual PDF editions on the IFU page; where no edition exists in a language, the English IFU applies. |
| Certificates | Provided to distributors and regulatory authorities on request. Registration status differs by country; confirm before importing.Source: company page. |
Where this page and the IFU differ, the IFU prevails. No figure on this page is a clinical outcome, and HANS Korea makes no claim of clinical superiority or equivalence to any other instrument.