The guideline's brush recommendation rests on trials of titanium brushes from other manufacturers. They are
listed here because they are what the guideline cites and because a dentist searching for evidence on
“titanium brushes” will meet them. The outcome lines repeat what each paper reports, without
comment. HANS Korea writes no manufacturer names here; none of these trials used the NiTi Brush, and they are
not on the Literature page.
-
Clinical evaluation of three surgical modalities in the treatment of peri-implantitis: a randomized controlled clinical trial
Toma S, Brecx MC, Lasserre JF · J Clin Med 2019;8(7):966 · cited by the guideline under R8.11
Reported. 47 patients were treated surgically with plastic curettes (15 patients, 25 implants), an air-abrasive device (16 patients, 22 implants) or a titanium brush (16 patients, 23 implants). At six months the authors report a greater reduction of gingival index and probing pocket depth in the titanium-brush group than in the other groups, and less bone loss than in the plastic-curette group. A successful outcome (implant present, PPD ≤ 5 mm, no bleeding on probing, no further mean bone loss ≥ 0.5 mm) was observed in 22% of implants with plastic curettes, 27% with the air-abrasive device and 33% with the titanium brush; the authors state that treatment success remained low.
The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.
DOI: 10.3390/jcm8070966
-
The adjunctive effect of a titanium brush in implant surface decontamination at peri-implantitis surgical regenerative interventions: a randomized controlled clinical trial
de Tapia B, Valles C, Ribeiro-Amaral T, Mor C, Herrera D, Sanz M, Nart J · J Clin Periodontol 2019;46(5):586–596 · cited by the guideline under R8.11
Reported. 30 patients, 15 per group, one-year follow-up. In both groups the implant surface was decontaminated with plastic ultrasonic scalers and 3% hydrogen peroxide and the intrabony defect was filled with an alloplastic material and covered with a collagen membrane; in the test group a titanium brush was applied in addition. At 12 months the reduction in probing pocket depth at the deepest site was 4.87 mm (SD 1.55) in the test group and 2.85 mm (SD 1.91) in the control group (p = 0.009); residual PPD was 3.6 mm and 4.92 mm respectively.
The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.
DOI: 10.1111/jcpe.13095
-
Implantoplasty vs. rotating titanium brushes in the surgical treatment of peri-implantitis: a 1-year randomised controlled clinical trial
Park SH, Kim DB, Kim DM, Sanz-Martin I, Sanz-Sanchez I, Derks J, Cha JK · J Clin Periodontol 2026;53(2):244–255 · published after the guideline; not cited by it
Reported. 30 patients, 15 implants per group, 12 months, non-inferiority design. Surface decontamination was performed with rotating titanium brushes (test) or implantoplasty (control). The composite outcome was achieved in 6 of 15 implants in the test group and 4 of 15 in the control group; probing pocket depth decreased by 3.6 ± 1.5 mm (test) and 3.3 ± 1.2 mm (control); bone levels changed by 0.0 ± 0.6 mm (test) and −0.7 ± 1.2 mm (control); time for surface decontamination was 3 min 1 s (test) and 5 min 27 s (control); patient-reported outcomes (OHIP-14) did not differ. The authors report that both groups showed favourable outcomes and that the pre-set non-inferiority criterion for the brush group was met.
The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.
DOI: 10.1111/jcpe.70056
No trial listed on this site, including these three, compared the NiTi Brush with another decontamination instrument. HANS Korea makes no claim of clinical superiority or equivalence.