What clinical guidelines say about titanium brushes in peri-implantitis surgery

This page is written for dentists and periodontists who want to know how the current clinical practice guideline treats rotary titanium brushes‡ as a step in the surgical treatment of peri-implantitis, and how the NiTi Brush relates to that guidance. It summarises one guideline whose text HANS Korea has read in full: the EFP S3 level clinical practice guideline on the prevention and treatment of peri-implant diseases (Herrera et al. 2023).1 The guideline does not name any manufacturer or product, and nothing on this page is a claim that the NiTi Brush is clinically superior or equivalent to any other instrument.

The guideline is published under a licence that does not permit reproduction of its text; its recommendations are restated here in HANS Korea's own words, with the recommendation numbers, grades and consensus levels as printed. Please read the original (DOI link in the references).

The EFP S3 guideline Herrera et al. 2023, J Clin Periodontol 50(Suppl 26):4–76

The European Federation of Periodontology (EFP) developed the guideline at the XVIII European Workshop in Periodontology (November 2022) following the methodology of the Association of Scientific Medical Societies in Germany (AWMF) and the GRADE approach. Thirteen commissioned systematic reviews provided the evidence; a panel of clinical experts and stakeholders voted on each recommendation. For the surgical step, the question of mechanical and physical implant-surface decontamination was answered by the systematic review of Ramanauskaite et al. 2023.2

The surgical procedure the guideline takes as its reference

In its introduction to the surgical step (section 8.1), the guideline describes the purpose of surgery as giving access to the implant so that the surface can be decontaminated, with resolution of the inflammatory lesion as the goal. The reference procedure it describes consists of flap elevation, removal of inflamed tissue, cleaning of the implant surface with, for example, small pieces of gauze soaked in saline, and removal of mineralised deposits with curettes. Reconstructive procedures, additional surface-decontamination methods and adjunctive antibiotics are listed as possible additions to that procedure.1

Recommendation 8.11 — mechanical and physical surface decontamination in surgery

ItemContent (restated)Grade
8.11 (1)The panel suggests not using air-polishing or Er:YAG laser for implant surface decontamination during the surgical treatment of peri-implantitis.Grade B, negative (“we suggest not”); consensus
8.11 (2)Titanium brushes may be considered as an alternative or adjunct to the reference decontamination (saline-soaked gauze and curettes).Grade O, open recommendation (“may be considered”); consensus, no abstentions
8.11 (3)The evidence is insufficient to make any recommendation on implantoplasty.Statement; consensus

Quality of evidence for the whole question: low. Source: Herrera et al. 2023, recommendation R8.11.1

What the evidence behind 8.11 (2) consisted of

The guideline's background text reports that the systematic review found five randomised controlled trials (183 patients, 242 implants, 6–24 months of follow-up) on mechanical or physical decontamination in surgery. Two of them studied a titanium brush: a three-arm trial in which the brush and glycine air-polishing were each compared with the reference instrumentation, and a trial in which a brush was added to ultrasonic instrumentation and hydrogen peroxide in regenerative surgery (both listed below). The guideline reports that a beneficial effect of the titanium brush was seen in those two trials, in bleeding on probing in one and in probing-depth reduction in the other, and grades the overall certainty of the evidence as low because the number of studies is small and they are heterogeneous. It adds that some decontamination protocols may be technically demanding and that some have not been tested for safety.1

Note on products. The guideline does not identify instruments by manufacturer. HANS Korea read the systematic review behind recommendation 8.11 (Ramanauskaite et al. 2023) in full: none of the five trials it includes names the NiTi Brush or HANS Korea, and none is on HANS Korea's Literature page, which lists only publications in which the instrument is identifiable as the NiTi Brush. The recommendation therefore applies to titanium brushes as a class of instrument, and it does not rest on any trial of the NiTi Brush.

The non-surgical step

For the non-surgical step the guideline recommends supra- and sub-marginal instrumentation with curettes and/or sonic or ultrasonic devices (recommendation R7.4, expert consensus, Grade A). The non-surgical recommendations and the systematic review behind them covered air-polishing, sonic and ultrasonic devices and lasers; they do not address rotary titanium brushes.1 This matches the position of the NiTi Brush: in the published studies it was used with surgical access to the implant surface, and its Instructions for Use describe mechanical debridement and cleaning of the contaminated implant surface in the dental office without defining a non-surgical application.3

Published trials on rotary titanium brushes as a class not the NiTi Brush

The guideline's brush recommendation rests on trials of titanium brushes from other manufacturers. They are listed here because they are what the guideline cites and because a dentist searching for evidence on “titanium brushes” will meet them. The outcome lines repeat what each paper reports, without comment. HANS Korea writes no manufacturer names here; none of these trials used the NiTi Brush, and they are not on the Literature page.

  • Clinical evaluation of three surgical modalities in the treatment of peri-implantitis: a randomized controlled clinical trial

    Reported. 47 patients were treated surgically with plastic curettes (15 patients, 25 implants), an air-abrasive device (16 patients, 22 implants) or a titanium brush (16 patients, 23 implants). At six months the authors report a greater reduction of gingival index and probing pocket depth in the titanium-brush group than in the other groups, and less bone loss than in the plastic-curette group. A successful outcome (implant present, PPD ≤ 5 mm, no bleeding on probing, no further mean bone loss ≥ 0.5 mm) was observed in 22% of implants with plastic curettes, 27% with the air-abrasive device and 33% with the titanium brush; the authors state that treatment success remained low.

    The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.

    DOI: 10.3390/jcm8070966

  • The adjunctive effect of a titanium brush in implant surface decontamination at peri-implantitis surgical regenerative interventions: a randomized controlled clinical trial

    Reported. 30 patients, 15 per group, one-year follow-up. In both groups the implant surface was decontaminated with plastic ultrasonic scalers and 3% hydrogen peroxide and the intrabony defect was filled with an alloplastic material and covered with a collagen membrane; in the test group a titanium brush was applied in addition. At 12 months the reduction in probing pocket depth at the deepest site was 4.87 mm (SD 1.55) in the test group and 2.85 mm (SD 1.91) in the control group (p = 0.009); residual PPD was 3.6 mm and 4.92 mm respectively.

    The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.

    DOI: 10.1111/jcpe.13095

  • Implantoplasty vs. rotating titanium brushes in the surgical treatment of peri-implantitis: a 1-year randomised controlled clinical trial

    Reported. 30 patients, 15 implants per group, 12 months, non-inferiority design. Surface decontamination was performed with rotating titanium brushes (test) or implantoplasty (control). The composite outcome was achieved in 6 of 15 implants in the test group and 4 of 15 in the control group; probing pocket depth decreased by 3.6 ± 1.5 mm (test) and 3.3 ± 1.2 mm (control); bone levels changed by 0.0 ± 0.6 mm (test) and −0.7 ± 1.2 mm (control); time for surface decontamination was 3 min 1 s (test) and 5 min 27 s (control); patient-reported outcomes (OHIP-14) did not differ. The authors report that both groups showed favourable outcomes and that the pre-set non-inferiority criterion for the brush group was met.

    The brush used in this trial was another manufacturer's titanium brush; the NiTi Brush was not used, and results obtained with one instrument cannot be transferred to another.

    DOI: 10.1111/jcpe.70056

No trial listed on this site, including these three, compared the NiTi Brush with another decontamination instrument. HANS Korea makes no claim of clinical superiority or equivalence.

Material constants, and two checks before any rotary brush is used

Rotary implant-surface brushes differ in bristle alloy. The figures below are published handbook values for the two alloys, as on the raw-material page. Material properties are not clinical outcomes; no clinical comparison is implied.

PropertyNickel-titanium (NiTi Brush bristle)Titanium
Modulus of elasticity (GPa)28–41115
Hardness (Vickers)303–362200–340
  • Check the speed. Check the rpm range the manufacturer specifies for the brush and the reduction ratio of your handpiece; the speed at the brush is the motor speed divided by that ratio. For the NiTi Brush the Instructions for Use specify a 20:1 reduction handpiece on an implant motor and 600–1,200 rpm.3
  • Check the registration. Check that the product is currently registered and supplied in your market, and read its Instructions for Use before the first case.

Safety: what the IFU says, and what the guideline reports about harms

The NiTi Brush

The Instructions for Use (HKIFU-001, Rev. 07) specify an implant motor with a 20:1 reduction handpiece at 600–1,200 rpm, extensive sterile saline irrigation throughout the brushing time, autoclave sterilisation before use (wrapped, 134 °C, 2.5 bar, 15 minutes, 20 minutes drying), single use, and a contraindication in patients who report a reaction to titanium or nickel.3 No instrument-related adverse events are reported in the trials listed on the Literature page, but those trials were not designed to collect them.

Harms reported in the guideline's background text (R8.11)

The guideline states that harms were not explicitly reported and evaluated in two of the five trials; that a slight pigmentation of the peri-implant soft tissue was observed in one of 30 patients treated with implantoplasty; that one trial reported adverse events in one of 16 patients, with persistence of suppuration and swelling, after air-polishing; and that one trial reported membrane exposure during healing after reconstructive therapy without giving the number affected. It adds that certain decontamination protocols have not been tested for safety.1 These are the guideline's observations on procedures; they are not presented here as advantages of any instrument.

Other guidance documents

HANS Korea checked, on 27 November 2026, which other guidance documents it could read in full and verify against the publisher's record. The 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions defines peri-implant health, mucositis and peri-implantitis (Berglundh et al. 2018); it is a classification, not a treatment guideline, and does not address decontamination instruments.4 Where a national society publishes its own guideline on peri-implantitis treatment, clinicians should consult that document; this page is limited to texts HANS Korea has read and can cite with a DOI, and will be extended as further guidelines are verified.

Where the NiTi Brush stands

The NiTi Brush is a rotary brush with nickel-titanium bristles, run from an implant motor through a 20:1 reduction handpiece at 600–1,200 rpm under sterile saline irrigation (Instructions for Use HKIFU-001, Rev. 07).3 It belongs to the class of instrument that recommendation 8.11 (2) describes as an alternative or adjunct to the reference decontamination.

In the randomised controlled trials on HANS Korea's Literature page, the NiTi Brush was the mechanical decontamination step of the surgical protocol, common to the treatment arms (in one trial, used in the flap-surgery arm): the multicentre trial of Derks et al. 2022 and its 3-year report (Alibegovic et al. 2025), the multicentre trial of Monje et al. 2025, the trials of Regidor et al. 2023 and 2026, Monje et al. 2023, and Soliman et al. 2026.5–11 Those trials evaluated other measures — bone substitutes, membranes, implantoplasty, adjunctive electrolytic or chemical decontamination — and none of them compared the brush with another decontamination instrument. The outcomes they report belong to the whole surgical protocol, not to the brush alone. How the patient and implant figures are counted is explained on the Literature summary page.

Do the guideline's citations and HANS Korea's list overlap?

HANS Korea checked this on 27 November 2026. Surface decontamination (recommendation 8.11): the five trials in the underlying review do not include any publication from HANS Korea's list. Reconstructive therapy (recommendations 8.7 and 8.8): the guideline reports seven randomised trials comparing reconstructive surgery with access flap surgery, from the review of Donos et al. 2023,12 but does not list them by name in its text, and the review's full text is not openly accessible; HANS Korea therefore does not state whether any of them used the NiTi Brush. The guideline's literature searches closed in 2022, before most of the NiTi Brush trials listed above were published. No part of the guideline's recommendations is presented here as resting on a trial of the NiTi Brush.

What this page does not say

It does not say that the guideline recommends the NiTi Brush, that a titanium brush is required, or that any brush gives better results than another method. The guideline grades the brush recommendation as open and the evidence as low. HANS Korea makes no claim of clinical superiority or equivalence. The choice of decontamination method is the clinician's decision, within the guideline applicable to them and the Instructions for Use of the device.

Questions clinicians ask

Does the EFP S3 guideline recommend titanium brushes for peri-implantitis?

It gives an open recommendation (Grade O): titanium brushes may be considered as an alternative or adjunct to the reference decontamination (saline-soaked gauze and curettes) during surgical treatment, with the quality of evidence graded low. It is not a positive recommendation in favour of brushes, and it names no product.1

Does the guideline recommend any specific decontamination method in surgery?

No method is recommended as superior. The guideline describes gauze soaked in saline plus curettes as its reference procedure, suggests not using air-polishing or Er:YAG laser for surface decontamination in surgery, leaves titanium brushes as an option, and makes no recommendation on implantoplasty.1

Was the NiTi Brush tested in the trials the guideline cites?

No. The guideline does not identify instruments by manufacturer, and none of the five trials in the review behind its brush recommendation names the NiTi Brush or appears on HANS Korea's Literature page. The NiTi Brush was the mechanical decontamination step in other randomised trials, listed on that page, which evaluated bone substitutes, membranes, implantoplasty or adjunctive decontamination and did not compare brushes with other instruments.5–11

What do trials of other manufacturers' titanium brushes show?

Three randomised trials report titanium brushes from other manufacturers: a three-arm trial against plastic curettes and an air-abrasive device (Toma et al. 2019), a trial adding a brush to ultrasonic and hydrogen-peroxide decontamination in regenerative surgery (de Tapia et al. 2019) and a trial against implantoplasty (Park et al. 2026). Their figures are repeated above as reported. The NiTi Brush was not used in them, and results obtained with one instrument cannot be transferred to another.13–15

Does the bristle alloy matter clinically?

Nickel-titanium has a modulus of elasticity of 28–41 GPa against 115 GPa for titanium, at Vickers hardness 303–362 against 200–340. These are material properties, not clinical outcomes; no trial has compared the two alloys clinically, and no clinical comparison is implied.

Can a titanium brush be used non-surgically?

The guideline's non-surgical recommendations call for curettes and/or sonic or ultrasonic instrumentation and do not address rotary brushes. In the published studies the NiTi Brush was used with surgical access, and its Instructions for Use do not define a non-surgical application.1,3

References

  1. Herrera D, Berglundh T, Schwarz F, Chapple I, Jepsen S, Sculean A, Kebschull M, Papapanou PN, Tonetti MS, Sanz M; EFP workshop participants and methodological consultant. Prevention and treatment of peri-implant diseases — The EFP S3 level clinical practice guideline. Journal of Clinical Periodontology 2023;50(Suppl 26):4–76. DOI: 10.1111/jcpe.13823
  2. Ramanauskaite A, Schwarz F, Cafferata EA, Sahrmann P. Photo/mechanical and physical implant surface decontamination approaches in conjunction with surgical peri-implantitis treatment: a systematic review. Journal of Clinical Periodontology 2023;50(Suppl 26):317–335. DOI: 10.1111/jcpe.13783
  3. Donos N, Calciolari E, Ghuman M, Baccini M, Sousa V, Nibali L. The efficacy of bone reconstructive therapies in the management of peri-implantitis. A systematic review and meta-analysis. Journal of Clinical Periodontology 2023;50(Suppl 26):285–316. DOI: 10.1111/jcpe.13775
  4. Toma S, Brecx MC, Lasserre JF. Clinical evaluation of three surgical modalities in the treatment of peri-implantitis: a randomized controlled clinical trial. Journal of Clinical Medicine 2019;8(7):966. DOI: 10.3390/jcm8070966
  5. de Tapia B, Valles C, Ribeiro-Amaral T, Mor C, Herrera D, Sanz M, Nart J. The adjunctive effect of a titanium brush in implant surface decontamination at peri-implantitis surgical regenerative interventions: a randomized controlled clinical trial. Journal of Clinical Periodontology 2019;46(5):586–596. DOI: 10.1111/jcpe.13095
  6. Park SH, Kim DB, Kim DM, Sanz-Martin I, Sanz-Sanchez I, Derks J, Cha JK. Implantoplasty vs. rotating titanium brushes in the surgical treatment of peri-implantitis: a 1-year randomised controlled clinical trial. Journal of Clinical Periodontology 2026;53(2):244–255. DOI: 10.1111/jcpe.70056
  7. HANS KOREA CO.,Ltd. NiTi Brush Instructions for Use, document HKIFU-001, Rev. 07 (22 May 2023). Download (PDF, multilingual)
  8. Berglundh T, Armitage G, Araujo MG, Avila-Ortiz G, Blanco J, Camargo PM, Chen S, Cochran D, Derks J, Figuero E, Hämmerle CHF, Heitz-Mayfield LJA, Huynh-Ba G, Iacono V, Koo KT, Lambert F, McCauley L, Quirynen M, Renvert S, Salvi GE, Schwarz F, Tarnow D, Tomasi C, Wang HL, Zitzmann N. Peri-implant diseases and conditions: consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions. Journal of Clinical Periodontology 2018;45(Suppl 20):S286–S291. DOI: 10.1111/jcpe.12957
  9. Derks J, Ortiz-Vigón A, Guerrero A, Donati M, Bressan E, Ghensi P, Schaller D, Tomasi C, Karlsson K, Abrahamsson I, Ichioka Y, Dionigi C, Regidor E, Berglundh T. Reconstructive surgical therapy of peri-implantitis: a multicenter randomized controlled clinical trial. Clinical Oral Implants Research 2022. DOI: 10.1111/clr.13972
  10. Alibegovic L, Trullenque-Eriksson A, Ortiz-Vigón A, Guerrero A, Donati M, Bressan E, Karlsson K, Dionigi C, Regidor E, Ichioka Y, Tomasi C, Ghensi P, Schaller D, Abrahamsson I, Berglundh T, Derks J. Bone substitute material in the surgical therapy of peri-implantitis: 3-year outcomes of a randomized controlled trial. Clinical Oral Implants Research 2025. DOI: 10.1111/clr.14393
  11. Monje A, Navarro-Mesa S, Soldini C, Zappalá G, Peña P, Navarro JM Sr, Pons R. Surface decontamination on the reconstructive therapy of peri-implantitis: a multicenter randomized clinical trial. Clinical Implant Dentistry and Related Research 2025. DOI: 10.1111/cid.70075
  12. Regidor E, Ortiz-Vigón A, Romandini M, Dionigi C, Derks J, Sanz M. The adjunctive effect of a resorbable membrane to a xenogeneic bone replacement graft in the reconstructive surgical therapy of peri-implantitis: a randomized clinical trial. Journal of Clinical Periodontology 2023. DOI: 10.1111/jcpe.13796
  13. Regidor E, Ortiz-Vigón A, Berglundh-Gottlieb J, Bougas K, Derks J. Adjunctive electrolytic implant surface decontamination in the reconstructive surgical treatment of peri-implantitis — a randomized clinical trial. Clinical Oral Implants Research 2026. DOI: 10.1111/clr.70152
  14. Monje A, Pons R, Vilarrasa J, Nart J, Wang H-L. Significance of barrier membrane on the reconstructive therapy of peri-implantitis: a randomized controlled trial. Journal of Periodontology 2023. DOI: 10.1002/JPER.22-0511
  15. Soliman OH, Malpassi C, Alanazi S, Emam ANM, Yousief SA, Baldini N. Regenerative surgical treatment of peri-implantitis: a one-year randomized clinical trial comparing autogenous bone and xenograft. Journal of Osseointegration 2026. DOI: 10.23805/JO.2026.764

HANS Korea is not the publisher of these works, holds no copyright in them and has no commercial relationship with the journals or the EFP. Listing a publication does not imply endorsement by its authors or institutions. See the legal notice.

Note on “titanium brush”

‡ Titanium brush (Ti brush) is used on this site as a generic, descriptive term, as in the literature; it is not a product name. Investigators have described the NiTi Brush by HANS Korea in this way, for example “a rotating titanium brush (Nano NiTi Brush, HANS KOREA CO. Ltd)” in randomised trials (Derks et al. 2022; Regidor et al. 2023; Alibegovic et al. 2025; Regidor et al. 2026) and “a titanium brush (Hans Korea Co., Ltd.)” in clinical studies (Galarraga-Vinueza et al. 2020; Schwarz et al. 2023; Zhou et al. 2026; Dionigi et al. 2025; Babgi & Felemban 2024); a laboratory study abbreviates it “TiB” (Ichioka et al. 2023). Another laboratory study calls it a “Ti brush” (“NiTi Brush Omega, Hans Korea”; Kao et al. 2024). The bristles are a nickel-titanium (NiTi) alloy, not pure titanium; for patients who report a reaction to titanium or nickel, see the IFU.