This page brings together, in one place, the operating conditions printed in the Instructions for Use of the NiTi Brush (document HKIFU-001, Rev. 07, 22 May 2023) and the published human clinical studies in which the instrument is identified by name. It is written for clinicians who already use, or are evaluating, a rotary implant-surface brush for the surgical treatment of peri-implantitis. The IFU is the binding document; this page summarises it and does not replace it. Nothing here is a claim of clinical superiority or equivalence to any other instrument: in the studies listed, the brush was the mechanical decontamination step of a surgical protocol, and the outcomes belong to the whole protocol, not to the brush alone.
Quotations from the IFU are reproduced as printed in the English edition. Settings in the study table are those reported by each paper.
Operating parameters in the IFU HKIFU-001, Rev. 07 — four items a clinician checks before the first case
IFU: “Recommended handpiece for NiTi Brush is the CE marked rotating implant motor (handpiece). Any handpiece that has the deceleration rate of ‘20:1’ is recommended.” Not intended for a high-speed turbine or a straight handpiece.
Source: IFU, “How to connect between NiTi brush and Dental Handpiece”.
IFU: “Recommended RPM for operation is 600 RPM ~ 1,200 RPM.” This is the speed at the brush, after the reduction ratio of the handpiece.
Source: IFU, same section. Published trial settings: see the table below.
IFU: “Rinse extensively with sterile saline solution throughout the whole brushing time.” Sterile saline is listed in the pre-use checklist “for irrigation and cooling”.
Source: IFU, “How to Use” and “Checklist before Use”.
Supplied non-sterile. IFU: autoclave, wrapped, 134 °C, 2.5 bar, 15 minutes exposure, 20 minutes drying; “After sterilization, use the NiTi Brush immediately without re-packing or re-storage.” “NiTi Brush is for single-use only. Do not re-use the NiTi Brush.”
Source: IFU, “Sterilization Method & Condition” and “Warning”.
Pre- and intra-operative checklist
Items marked IFU are taken from the Instructions for Use. Items marked as reported describe where the brushing step sat in the published surgical protocols.
-
Confirm the speed at the brush, not the number on the motor display
IFU. 600–1,200 rpm through a 20:1 reduction handpiece on an implant motor. With a reduction handpiece in the chain, the motor display shows the motor speed; the speed at the brush is that figure divided by the reduction ratio. Check the ratio of the handpiece you own and set the motor so that the brush turns inside the IFU range.
As reported. The published trials that give a figure used up to 1,200 rpm under continuous irrigation (Derks et al. 2022; Regidor et al. 2023, 2026) or about 600 rpm for about 2–3 minutes (Monje et al. 2023, 2025, 2026). Study table ↓
-
Keep sterile saline irrigation running for the whole brushing time
IFU. “Rinse extensively with sterile saline solution throughout the whole brushing time.” Sterile saline is prepared before the case “for irrigation and cooling”. The IFU gives no further statement on temperature; the irrigation requirement is what it says.
-
Inspect the brush and use it once
IFU. Before use: “Visually inspect the damage, breakage, deformation, pollution from any foreign material in NiTi Brush”, and “In case of damage, breakage, deformation, pollution from any foreign material in NiTi Brush, do not use it.” Do not use the brush if the inner plastic box is broken or the expiry date has passed. After the case: “Because of the risk for cross infection from patient to patient, NiTi Brush should be discarded after use.” A used brush is not re-sterilised.
IFU, several implants in one patient. “In case of multiple use in one patient using one NiTi Brush, it is recommended to immerse the NiTi Brush in a 3% hydrogen peroxide solution and rinse with sterile saline before clean or debride another implant.”
-
Check the patient history for reactions to titanium or nickel
IFU. “NiTi Brush must not be used in patients reporting reactions against or incompatibility with titanium or nickel.” The pre-use checklist also asks for the operating-room temperature (15–30 °C) and the recommended handpiece to be ready.
-
Brushing is one step of the surgery, not the treatment
As reported. In the published surgical protocols the brush was used after flap reflection and removal of granulation tissue (for example Zhou et al. 2026: “After removal of granulation tissues with a curette, implant surfaces were decontaminated using a titanium brush‡”), followed by the resective or reconstructive measures the authors had chosen — bone substitutes, membranes, implantoplasty, adjunctive chemical or electrolytic steps. None of the listed trials compared one decontamination instrument with another. Surgical protocol as published → · Chemical adjuncts after brushing →
Clinician FAQ answers are on the page; open an item to read it
What are the operating conditions in the Instructions for Use?
The IFU (HKIFU-001, Rev. 07) recommends a CE-marked rotating implant motor with a 20:1 reduction handpiece and a speed of 600 to 1,200 rpm at the brush, with extensive sterile saline irrigation throughout the whole brushing time. The brush is supplied non-sterile, is autoclaved wrapped at 134 °C and 2.5 bar for 15 minutes with 20 minutes drying before use, is used immediately after sterilisation, and is single use. It must not be used in patients who report reactions to titanium or nickel.
Which speed and brushing time did the published multicentre trials report?
Two settings appear in the trials that name the instrument. The multicentre randomised trial of Derks et al. 2022 and the trials of Regidor et al. 2023 and 2026 report the brush at up to 1,200 rpm under continuous saline irrigation. The multicentre randomised trial of Monje et al. 2025 and the trials of Monje et al. 2023 and 2026 report “600 rpm for approximately 2–3 min”. Both settings lie inside the IFU range. The brushing time is what the authors report; the IFU does not prescribe a duration, and HANS Korea’s protocol page notes that it depends on the defect and the surface.
How is the volume of evidence counted?
A paper is counted only when it has been read in full, verified against the publisher’s record and identifies the instrument as the NiTi Brush of HANS Korea (under its current name, its former name I.C.T., or the RotoBrush name of the United States distributor). Papers that say only “a titanium brush” are excluded. On that rule the Literature page lists 34 verified publications. Patient and implant totals are taken from the clinical studies that report them and count each cohort once: a 3-year report, a secondary analysis or a single-centre report of a trial already counted is not added again, and in the one trial where the brush was used in one arm only, only that arm is counted. The totals so counted are 723 patients and at least 1,088 implants, followed 6 to 120 months. The counting rules and the study-by-study table are on the Literature summary page.
Does any listed study compare the NiTi Brush with another instrument?
No. In the randomised trials listed, the NiTi Brush was the mechanical decontamination step common to the treatment arms (in one trial, the flap-surgery arm), and the trials evaluated other measures: bone substitutes, membranes, implantoplasty, adjunctive chemical or electrolytic decontamination. The outcomes reported belong to the whole surgical protocol. HANS Korea makes no claim of clinical superiority or equivalence to any other instrument.
Where is the full IFU?
The current Instructions for Use (HKIFU-001, Rev. 07) can be downloaded from the IFU page as PDF in the languages listed there. The Korean IFU is available to registered Korean dentists on the Korean pages. Where no edition exists in a language, the English IFU applies. Regulatory clearance and availability differ by country; confirm the status for your market with HANS Korea or the local authorised distributor.
Notice to authors, publishers and manufacturers
The Literature pages summarise published research so that clinicians can find and read the original papers. HANS Korea has no intention of infringing anyone’s copyright, trademark or other intellectual-property rights: quotations are brief and attributed, every entry links to the publisher’s own page, no full text or figure is reproduced, and no comparative claim is made about any other manufacturer’s product. If you are an author, publisher or manufacturer and have any concern about how your work or product is referred to on this site, please contact [email protected]. We will review it promptly and correct or remove the material. Conversely, the summaries, tables, compiled figures, images and documents prepared by HANS Korea for this site are its own work: their use by any third party — in particular by a manufacturer, importer or seller of a similar device — for promotion, product development or any other purpose requires the prior written consent of HANS KOREA CO.,Ltd.
Human clinical studies naming the NiTi Brush as listed on the Literature page; cohorts counted once
The table lists the clinical studies from the Literature page that report patients: randomised and controlled trials first, then cohort studies, case series and case reports. Translational and laboratory studies and reviews are on the Literature page but not here. “Reported brush protocol” gives the setting each paper reports; wording in quotation marks is quoted from papers published under a free licence, the rest is stated as fact. Where a paper names the manufacturer but gives no settings, the cell says so. Rows in grey are follow-up reports, secondary analyses or subsets of a cohort already listed and are not added to the totals of 723 patients and at least 1,088 implants. Every DOI link goes to the publisher’s landing page.
| Authors & journal | Study design | Patients / implants · follow-up | Reported brush protocol | DOI |
|---|---|---|---|---|
| Derks et al. 2022 Clin Oral Implants Res | Multicentre randomised controlled trial | 138 / 147 · 12 months | Nano NiTi Brush; up to 1,200 rpm; continuous irrigation | 10.1111/clr.13972 |
| Alibegovic et al. 2025 Clin Oral Implants Res | 3-year report of the Derks et al. 2022 trial | 120 / 127 · 36 months | Nano NiTi Brush; saline irrigation (settings as in the 2022 report) | 10.1111/clr.14393 |
| Monje et al. 2025 Clin Implant Dent Relat Res | Multicentre randomised clinical trial (surface decontamination) | 58 / ≥ 58 · 12 months | “NiTi brushes (Hans Korea Co., Gyeonggi-do, Korea) at 600 rpm for approximately 2–3 min” | 10.1111/cid.70075 |
| Monje, Pons, Peña 2024 Int J Periodontics Restorative Dent | Single-centre report of the Monje et al. 2025 trial | 19 / 23 · 12 months | NiTi Brush; 600 rpm; about 2–3 minutes | 10.11607/prd.7151 |
| Regidor et al. 2023 J Clin Periodontol | Randomised clinical trial (resorbable membrane) | 43 / 43 · 12 months | Nano NiTi Brush; up to 1,200 rpm; continuous saline irrigation | 10.1111/jcpe.13796 |
| Regidor et al. 2026 Clin Oral Implants Res | Randomised clinical trial (adjunctive electrolytic cleaning) | 40 / 44 · 12 months | Nano NiTi Brush; up to 1,200 rpm; continuous saline irrigation | 10.1111/clr.70152 |
| Malmqvist et al. 2025 J Periodontol | Randomised controlled trial; brush used in the flap-surgery arm only | 12 / 12 (flap-surgery arm) · 6 months | “the exposed threads of the implants were brushed with a titanium brush (NiTi Brush Pocket, HANS KOREA Co. Ltd., Goyang, South Korea)” — no speed stated | 10.1002/JPER.24-0683 |
| Monje et al. 2023 J Periodontol | Randomised controlled trial (barrier membrane) | 33 / 48 · 12 months | NiTi brushes (Hans Korea Co., Gyeonggi-do, Korea) for about 2–3 min at 600 rpm | 10.1002/JPER.22-0511 |
| Monje et al. 2023 Clin Implant Dent Relat Res | Secondary analysis (defect angle) of the Monje et al. 2023 trial | 33 / 48 · 12 months | NiTi brushes (Hans Korea Co., Gyeonggi-do, Korea) — no separate settings reported | 10.1111/cid.13244 |
| Monje et al. 2026 Clin Implant Dent Relat Res | Quasi-randomised clinical trial (extent of implantoplasty) | 33 / 40 · 12 months | “NiTi brushes (Hans Korea Co., Gyeonggi-do, Korea) at 600 rpm for approximately 2–3 min”, in both groups | 10.1111/cid.70144 |
| Soliman et al. 2026 J Osseointegr | Split-mouth randomised clinical trial (autogenous bone vs xenograft) | 10 / 20 · 12 months | “mechanical debridement of the implant surface with titanium brushes (NiTi-brush, Hans Korea Co Ltd) under copious sterile saline irrigation”, in both groups | 10.23805/JO.2026.764 |
| Monje et al. 2025 Int J Oral Implantol | Case series (rhPDGF-BB reconstructive therapy) | 10 / 13 · 12 months | NiTi Brush; 600 rpm; about 2–3 minutes | PMID 40047362 (no DOI) |
| Monje et al. 2021 J Periodontol | Retrospective study (implantoplasty) | 43 / 135 · about 24 months | NiTi Brush; no settings reported | 10.1002/JPER.21-0103 |
| Zhou et al. 2026 J Clin Periodontol | Retrospective cohort, university centre, 2018–2023 | 223 / 406 · mean 30.8 months | “After removal of granulation tissues with a curette, implant surfaces were decontaminated using a titanium brush (Hans Korea Co. Ltd.)” | 10.1111/jcpe.70115 |
| Schwarz et al. 2023 Clin Oral Implants Res | Retrospective analysis, same centre, 2018–2021 | 36 / 40 · 12 months | Titanium brush of HANS Korea; no settings reported | 10.1111/clr.14057 |
| Galarraga-Vinueza et al. 2020 J Clin Periodontol | Pilot study | 20 / 28 · 6 months | Titanium brush of HANS Korea; no settings reported | 10.1111/jcpe.13335 |
| Brincat et al. 2023 Appl Sci | Case series (implantoplasty) | 57 / 91 · 12–42 months | “NiTibrushes (ICT Micro and ICT Nano, Hans Korea® Gyeonggi-do, South Korea)” | 10.3390/app13127224 |
| Babgi & Felemban 2024 Umm Al-Qura University | Report of two cases | 2 / 2 · — | NiTi Brush OMEGA on a low-speed handpiece; no speed stated | 10.54940/ms75920247 |
| Parma-Benfenati et al. 2026 Clin Adv Periodontics | 10-year follow-up case report | 1 / 1 · 120 months | “final mechanical debridement was performed with a titanium brush (NiTi Brush, HANS Korea Co., Ltd.)” | 10.1002/cap.10359 |
Patient and implant figures and the counting rule are those of the Literature summary page (Revision 2, 12 October 2026); the quoted instrument wording is that of the Literature page. Listing a publication does not imply endorsement by its authors or institutions. Outcome data are deliberately not repeated here; please read the original publication.
Background: what a titanium brush is and the titanium brush for peri-implantitis (NiTi Brush models and evidence).
Scope of this page
The studies listed are surgical protocols in which the NiTi Brush was the mechanical implant-surface decontamination step, alongside flap management, removal of granulation tissue and the resective or regenerative procedures chosen by the authors. The outcomes they report belong to those protocols as a whole. This page is professional information for dental clinicians; it is not medical advice and does not replace the Instructions for Use. Before use, clinicians should verify the regulatory clearance of the device in their country, the compatibility of their implant motor and reduction handpiece with the IFU, and the current IFU edition for their region. Regulatory clearance and availability differ by country; not every model is registered in every market.
Note on “titanium brush”
‡ Titanium brush (Ti brush) is used on this site as a generic, descriptive term, as in the literature; it is not a product name. Investigators have described the NiTi Brush by HANS Korea in this way, for example “a rotating titanium brush (Nano NiTi Brush, HANS KOREA CO. Ltd)” in randomised trials (Derks et al. 2022; Regidor et al. 2023; Alibegovic et al. 2025; Regidor et al. 2026) and “a titanium brush (Hans Korea Co., Ltd.)” in clinical studies (Galarraga-Vinueza et al. 2020; Schwarz et al. 2023; Zhou et al. 2026; Dionigi et al. 2025; Babgi & Felemban 2024); a laboratory study abbreviates it “TiB” (Ichioka et al. 2023). Another laboratory study calls it a “Ti brush” (“NiTi Brush Omega, Hans Korea”; Kao et al. 2024). The bristles are a nickel-titanium (NiTi) alloy, not pure titanium; for patients who report a reaction to titanium or nickel, see the IFU.