NiTi Brush — safety, contraindications and limits

This page collects, in one place, what the Instructions for Use of the NiTi Brush (document HKIFU-001, Rev. 07, 22 May 2023) say about contraindications, warnings and handling, and what the published laboratory studies that name the instrument report about its limits: titanium particle release, biofilm left behind, and change to the implant surface. It is written for clinicians who are deciding whether and how to use a rotary implant-surface brush. The IFU is the binding document; this page summarises it and does not replace it. Where the IFU is silent on a point, the page says so rather than adding advice.

IFU wording is reproduced as printed in the English edition. Laboratory figures are those printed on the Literature page and in the papers cited.

Contraindications and warnings IFU HKIFU-001, Rev. 07

ItemWhat the IFU says
Titanium or nickel hypersensitivityWarning: “NiTi Brush must not be used in patients reporting reactions against or incompatibility with titanium or nickel.” The pre-use checklist lists “The reactions of patient against or incompatibility with titanium or nickel.” The bristles are a nickel-titanium alloy.
Other contraindications“It is contraindicated in young patient with poor attention and it can’t be used in non-implant surgery patients.” For pregnant, paediatric or elderly patients: “there is no safety information, please consult to specialist.”
Side effects listed“Redness, Discomfort, Swelling.”
Patient populationAdults with dental implants.
Intended operator“Only use by dentist or dental hygienist and trained dental assistants.”
Intended use“mechanical debridement and cleaning of titanium dental implant surface contaminated by osseous defect of plague, calculus or any other foreign body materials resulted from peri-implantitis”, in the dental office. Indication: peri-implantitis.

Handling: single use, sterilisation, inspection

  1. Supplied non-sterile; sterilise before use

    IFU: “NiTi Brush is non-sterile product.” Autoclave, wrapped, 134 °C, 2.5 bar, 15 minutes exposure, 20 minutes drying. “After sterilization, use the NiTi Brush immediately without re-packing or re-storage.” Keep the brush in the sterilisation pouch until use.

  2. Single use

    IFU: “NiTi Brush is for single-use only. Do not re-use the NiTi Brush.” “Because of the risk for cross infection from patient to patient, NiTi Brush should be discarded after use.” Disposal follows the local rules for disposable medical devices.

  3. Several implants in one patient

    IFU: “In case of multiple use in one patient using one NiTi Brush, it is recommended to immerse the NiTi Brush in a 3% hydrogen peroxide solution and rinse with sterile saline before clean or debride another implant.”

  4. Inspect before use; do not use a damaged brush

    IFU: “Visually inspect the damage, breakage, deformation, pollution from any foreign material in NiTi Brush” and “In case of damage, breakage, deformation, pollution from any foreign material in NiTi Brush, do not use it.” Do not use if the inner plastic box is broken or the expiry date (5 years from manufacture) has passed.

  5. Drive and irrigation

    IFU: CE-marked implant motor, handpiece with a 20:1 reduction ratio, 600–1,200 rpm; “Rinse extensively with sterile saline solution throughout the whole brushing time.” Operating-room temperature 15–30 °C; storage 15–30 °C, 1–75% humidity, away from sunlight.

What the IFU does not address

The IFU gives no instruction on bristle loss or bristle fracture during use beyond the inspection rule above (do not use a damaged, broken or deformed brush), no brushing time, no statement on pressure or angulation, and no instruction for use without surgical access. HANS Korea does not add technique advice here; where the IFU is silent, the clinician's judgement and the published protocols apply.

Non-surgical use is not described in the IFU

The intended use in the IFU is mechanical debridement and cleaning of the contaminated implant surface; the IFU does not describe a non-surgical (closed) application. In the published clinical studies on the Literature page, the brush was used on the exposed implant surface after flap elevation and removal of granulation tissue, as the mechanical decontamination step of a surgical protocol. The current EFP S3 guideline recommends curettes and/or sonic or ultrasonic instrumentation for the non-surgical step and does not address rotary brushes there (see what clinical guidelines say).

Limits reported in laboratory studies naming the instrument

In vitro findings describe the instrument on discs or model implants, not clinical outcomes. They are reported here as the papers state them, including the unfavourable ones.

  • Titanium particle release

    The NiTi Brush (OMEGA model) released 6.45 ± 2.03 µg of titanium per disc, about one-fifth of the amount released by metal-tip ultrasonic scaling (34.0–34.4 µg); erythritol air-polishing and a PEEK-tip ultrasonic released less than the brush. Rotary brushing therefore does release titanium particles.1

    Related clinical observation: in peri-implant soft tissue from patients whose implants had been decontaminated with titanium curettes and a rotating titanium brush (NANO model), titanium micro-particles were a common finding and were not associated with the occurrence of peri-implantitis (Dionigi et al. 2025).2

  • Biofilm is not completely removed

    Protocols that included the rotating brush (1,200 rpm under saline) or air-polishing removed more biofilm and surface carbon than gauze wiping alone; no protocol removed the biofilm completely.3 In an earlier disc study by the same group, residual bacteria were close to zero on smooth discs with every method tested while some remained on shot-blasted discs, and no method restored pristine cell compatibility (Ichioka et al. 2022).4

  • The implant surface is changed

    On clean RBM and SLA titanium discs the brush (former name I.C.T., 800 rpm) reduced surface roughness (Ra 1.43 to 1.15 µm on RBM, 2.85 to 2.23 µm on SLA).5 On uncontaminated discs brushed at 400 and 600 rpm, brushing left visible grooves and a slight increase in roughness on titanium and no significant change on a zirconia-coated surface; dental-pulp stem cells kept their biological features after either cleaning procedure.6 The brush is intended to remove the contaminated layer; the papers show that it also alters the micro-topography of the surface it touches.

Outcomes belong to the whole protocol

In the clinical studies that name the NiTi Brush, it was the mechanical decontamination step of a surgical protocol that also included flap management, removal of granulation tissue, in most cases a chemical step, and the resective or reconstructive procedure chosen by the authors. Their outcomes belong to that whole protocol, not to the brush alone, and no study compared the brush with another decontamination instrument. No instrument-related adverse events are reported in those trials, but they were not designed to collect them. HANS Korea makes no claim of clinical superiority or equivalence to any other instrument.

Questions clinicians ask

Can the NiTi Brush be used in a patient with a nickel allergy?

No. The IFU states that the brush must not be used in patients reporting reactions against, or incompatibility with, titanium or nickel. The bristles are a nickel-titanium alloy. Ask before the case; it is an item on the IFU's pre-use checklist.

Can a NiTi Brush be re-sterilised and used on another patient?

No. The IFU specifies single use and discarding after use because of the risk of cross-infection. Within one patient, when several implants are treated with one brush, the IFU recommends immersion in 3% hydrogen peroxide and a sterile-saline rinse between implants.

What should be done if a bristle breaks or is lost during use?

The IFU does not address this. It requires the brush to be inspected before use and not used if damaged, broken or deformed. HANS Korea gives no further instruction on this page; management of any retained fragment is the clinician's decision.

Does the brush release titanium particles?

Yes. In vitro, the NiTi Brush released 6.45 ± 2.03 µg of titanium per disc, about one-fifth of metal-tip ultrasonic scaling; erythritol air-polishing and a PEEK ultrasonic tip released less (Kao et al. 2024). In peri-implant soft tissue of treated patients, titanium micro-particles were common and not associated with peri-implantitis (Dionigi et al. 2025).

Does the brush remove all the biofilm?

Not in vitro. On model implants with a multispecies biofilm, protocols with the rotating brush removed more biofilm than gauze wiping alone, but no protocol removed it completely (Ichioka et al. 2023).

Can it be used without raising a flap?

The IFU does not describe a non-surgical application, and in the published studies the brush was used on the exposed surface after flap elevation. The current EFP S3 guideline recommends curettes and/or sonic or ultrasonic instrumentation for the non-surgical step.

References

  1. Kao A, Tawse-Smith A, Ma S, Duncan WJ, Reid M, Atieh MA. Quantity and size of titanium particles released from different mechanical decontamination procedures on titanium discs: an in vitro study. Dentistry Journal 2024;12(5):123. DOI: 10.3390/dj12050123
  2. Dionigi C, Nagy G, Derks J, Ichioka Y, Tomasi C, Larsson L, Primetzhofer D, Berglundh T. Titanium micro-particles are commonly found in soft tissues surrounding dental implants. Communications Medicine 2025. DOI: 10.1038/s43856-025-00756-3
  3. Ichioka Y, Virto L, Nuevo P, Gamonal JD, Derks J, Larsson L, Sanz M, Berglundh T. Decontamination of biofilm-contaminated implant surfaces: an in vitro evaluation. Clinical Oral Implants Research 2023;34(10):1058–1072. DOI: 10.1111/clr.14136
  4. Ichioka Y, Derks J, Dahlén G, Berglundh T, Larsson L. Mechanical removal of biofilm on titanium discs: an in vitro study. Journal of Biomedical Materials Research Part B: Applied Biomaterials 2022;110(5):1044–1055. DOI: 10.1002/jbm.b.34978
  5. Kim SK, Lee JK, Park SH, Chang BS, Lee SY, Um HS. Altered bacterial adhesion with changes in roughness of titanium surfaces after implant cleaning. Oral Biology Research 2016;40(1):9–15. DOI: 10.21851/obr.40.1.201603.9
  6. Tagliaferri N, Pisciotta A, Orlandi G, Bertani G, Di Tinco R, Bertoni L, Sena P, Lunghi A, Bianchi M, Veneri F, Bellini P, Bertacchini J, Conserva E, Consolo U, Carnevale G. Zirconia hybrid dental implants influence the biological properties of neural crest-derived mesenchymal stromal cells. Nanomaterials 2024;14(5):392. DOI: 10.3390/nano14050392
  7. HANS KOREA CO.,Ltd. NiTi Brush Instructions for Use, document HKIFU-001, Rev. 07 (22 May 2023). Download (PDF, multilingual)