Implant surface decontamination in peri-implantitis: mechanical debridement and the other methods

Implant surface decontamination is the step of peri-implantitis treatment in which the biofilm and deposits on the exposed implant surface are removed, so that the inflamed tissue can heal. It is done mechanically (mechanical debridement with curettes, ultrasonic or sonic tips, air-polishing or rotary titanium brushes‡), by reshaping the surface (implantoplasty), with chemical agents, or with physical methods such as lasers, photodynamic therapy, electrolytic cleaning or cold plasma — usually several of them combined. This page explains each class of method and what the current clinical practice guidelines say about it.

Written by HANS KOREA CO.,Ltd., the manufacturer of the NiTi Brush, for dentists. It describes methods as classes and names no other manufacturer’s products; no method is ranked against another. The guidelines are restated in our own words with their recommendation numbers and grades — please read the originals (references below).

Four questions first

Literature basis: Literature Rev. 4 · 2026-11-29 · version history; how the figures are counted: counting rules.

How do mechanical and chemical approaches differ in implant surface decontamination?

Mechanical debridement physically removes biofilm and deposits from the threads and the rough surface of the exposed implant. Chemical agents are applied afterwards as an adjunct, to reduce or inactivate the bacteria that remain. The EFP S3 guideline takes gauze soaked in saline plus curettes as its reference decontamination in surgery and states that titanium brushes may be considered as an alternative or adjunct to it; published surgical protocols usually combine a mechanical and a chemical step.1

In which studies were titanium brushes and the NiTi Brush used?

The publications in which the instrument is identifiable as the NiTi Brush are listed, with how they are counted, on the Published studies page. In the randomised trials among them the brush was, with one exception, a step common to both arms: the trials compared something else — a bone substitute, a barrier membrane, electrolytic cleaning, a chemical adjunct or the extent of implantoplasty. Randomised trials that added a brush to one arm or compared a brush directly with another instrument were performed with titanium brushes of other manufacturers; the NiTi Brush was used in one trial only in the flap-surgery arm, compared with diode-laser treatment (see the evidence table).

What did studies that added an adjunct after brushing compare?

They compared adjuncts with each other after the same mechanical cleaning with the NiTi Brush. In a multicentre randomised trial (58 patients) comparing electrolytic cleaning with hydrogen peroxide, disease resolution at 12 months was 87.5% with electrolytic cleaning and 64.5% with hydrogen peroxide; the difference was not statistically significant (p = 0.08). Both groups used the NiTi Brush for mechanical cleaning. Another randomised trial (40 patients) compared the brush alone with the brush plus electrolytic cleaning and found no significant difference between the groups. Laboratory studies compared the effect of brushes, a polishing brush and agents on surface morphology and cell compatibility.

What did studies in patients and laboratory studies each measure?

Studies in patients reported probing-depth reduction, radiographic bone change, disease resolution and implant survival as the result of the whole treatment protocol, so they cannot be read as the effect of the brush alone. Laboratory studies measured biofilm removal, surface roughness and morphology, titanium particle release and cell response on discs or models; no method removed all biofilm, and brushes were reported to alter micro-rough surfaces. Laboratory results do not show clinical effect directly.

Why the implant surface has to be decontaminated

Peri-implantitis is a plaque-associated inflammatory condition of the tissues around a dental implant with loss of supporting bone.3 The biofilm that sustains it sits on the implant itself: on the threads and on the micro-rough surface that most implants carry. Unlike a natural root, that surface cannot be planed smooth without changing the implant, and the threads leave recesses that a flat instrument does not reach. Decontamination therefore aims to remove biofilm and deposits from the threads and the rough surface while keeping the implant’s geometry, or — in implantoplasty — deliberately removes the threads and the rough surface instead.

In non-surgical treatment the surface is reached through the pocket. In surgical treatment a flap is raised, the inflamed tissue is removed and the surface is cleaned under direct vision; the EFP S3 guideline takes as its reference surgical procedure flap elevation, removal of inflamed tissue, cleaning of the surface with, for example, small pieces of gauze soaked in saline, and removal of mineralised deposits with curettes.1

The methods, class by class

MethodWhat it doesWhere it is usedGuideline position (summary)
Curettes and hand instrumentsPlastic, carbon-fibre, titanium or steel blades scrape deposits off the surface.Non-surgical and surgicalEFP: part of the reference procedure (gauze soaked in saline + curettes) and of non-surgical instrumentation (R7.4).1
Ultrasonic and sonic tipsVibrating tips, often with plastic or carbon sleeves, loosen deposits under irrigation.Non-surgical and surgicalEFP: non-surgical instrumentation with curettes and/or sonic or ultrasonic devices (R7.4).1
Air-polishingA jet of fine powder (glycine or erythritol), air and water removes biofilm.Non-surgical and surgicalEFP: suggested not to be used for surface decontamination during surgery (R8.11 (1), Grade B).1 German S3: evidence for glycine air-polishing in non-reconstructive surgery (Rec. 18, Grade B).2
Rotary titanium brushesA small rotary brush with titanium-alloy bristles, run in an implant motor and handpiece under saline, reaches into the threads.Surgical (with access to the surface)EFP: may be considered as an alternative or adjunct to the reference decontamination (R8.11 (2), Grade O, evidence low).1 German S3: evidence for a titanium brush in non-reconstructive surgery (Rec. 18, Grade B); as an adjunct in combined surgery it can be dispensed with (Rec. 20, Grade 0).2
ImplantoplastyBurs and polishers remove the exposed threads and the rough surface, leaving a smooth, narrower implant.Surgical, usually non-reconstructiveEFP: evidence insufficient for any recommendation (R8.11 (3)).1
Chemical agentsSolutions such as saline, hydrogen peroxide, chlorhexidine, citric acid or tetracycline are applied to the surface, usually after mechanical cleaning.Mostly surgical, as an adjunctEFP: chlorhexidine and photodynamic therapy suggested not to be used for surface decontamination during surgery (R8.12, Grade B, evidence very low); in non-surgical therapy, a desiccant antiseptic gel and locally administered antimicrobials suggested not to be used (R7.8, R7.9, Grade B).1 German S3: adjunctive antiseptics should not be used as part of the decontamination in non-reconstructive surgery (Rec. 18, Grade B).2 Neither guideline makes a specific recommendation on tetracycline solutions on the implant surface. Studied as adjuncts to mechanical cleaning; see the reader chapter for the papers.
Lasers and photodynamic therapyLaser energy (for example Er:YAG or diode) or a light-activated dye is used to remove or inactivate biofilm.Non-surgical and surgicalEFP: Er:YAG laser suggested not to be used for surface decontamination during surgery (R8.11 (1), Grade B).1 German S3: as an alternative in combined surgery it can be dispensed with (Rec. 20, Grade 0).2
Electrolytic cleaning and cold plasmaAn electric current through the implant in a salt solution, or ionised gas, is used to detach or inactivate biofilm.Surgical (electrolytic); experimental and surgical (plasma)Newer methods; see the reader chapters for the trials and laboratory studies.

Grades: in the EFP guideline Grade A = “we recommend”, B = “we suggest”, O = “may be considered”; in the German guideline B = “should”, 0 = “can”. The EFP guideline grades the quality of evidence for surgical surface decontamination as a whole as low.1

What the studies actually compared

Look first at what each study compared. Tags: Both arms: NiTi Brush the brush was a step common to all arms (its own effect is not isolated) · Direct comparison instruments or methods compared with each other · Laboratory discs, models or animals (does not show clinical effect directly) · Other maker's brush a titanium brush of another manufacturer · Guideline / review. Results are reported as the papers state them; they are not HANS KOREA’s assessment.

Representative studies of implant surface decontamination in peri-implantitis
MethodStudy typeInstrument / agentWhat was comparedResult and limitsDOI
Surgical surface decontamination overall
Guideline / review
Clinical practice guideline (EFP S3, 2023)Saline-soaked gauze, curettes, titanium brushes, air-polishing, lasers, implantoplasty, chemical agentsRecommendation grades by methodGauze and curettes as the reference; titanium brushes may be considered as an alternative or adjunct (R8.11, grade O); chlorhexidine and photodynamic therapy suggested not to be used for surface decontamination during surgery (R8.12, grade B). Overall quality of evidence low.10.1111/jcpe.13823
Non-reconstructive and combined surgery
Guideline / review
Clinical practice guideline (German S3, 2022)Titanium brush, glycine air-polishing, lasers, antisepticsRecommendation grades by methodEvidence for a titanium brush and glycine air-polishing in non-reconstructive surgery; adjunctive antiseptics should not be used there (Rec. 18, grade B). An adjunctive titanium brush can be dispensed with in combined surgery (Rec. 20, grade 0).AWMF 083-023
Mechanical and physical decontamination
Guideline / review
Systematic review (Ramanauskaite et al. 2023)Mechanical and physical methods during surgeryFive randomised trials of surgical decontaminationData limited, clinical efficacy inconclusive; the authors state that titanium brushes may be beneficial in reducing signs of inflammation.10.1111/jcpe.13783
Reconstructive surgery + bone substitute
Both arms: NiTi Brush
Multicentre RCT (Derks et al. 2022), 138 patients / 147 implantsNiTi Brush NANO, ≤1,200 rpm, saline irrigationBone substitute vs noneAt 12 months probing depth reduced by about 3.7 mm and bone gained about 1.0 mm (whole treatment); the brush’s own effect is not isolated.10.1111/clr.13972
The same trial at 3 years
Both arms: NiTi Brush
3-year follow-up of the RCT (Alibegovic et al. 2025)NiTi Brush NANOBone substitute vs noneAt 3 years probing depth reduced by 3.2–3.5 mm; 14 of 127 implants (11%) lost; disease resolution 39%. Shows the limits of long-term outcomes.10.1111/clr.14393
Reconstructive surgery + membrane
Both arms: NiTi Brush
RCT (Regidor et al. 2023), 43 patientsNiTi Brush NANO, ≤1,200 rpmResorbable membrane vs noneNo implant lost at 12 months; the comparison was the membrane.10.1111/jcpe.13796
Reconstructive surgery + membrane
Both arms: NiTi Brush
RCT (Monje et al. 2023), 33 patients / 48 implantsNiTi Brush, 600 rpm, about 2–3 minBarrier membrane vs noneMean disease resolution 77.1% at 12 months (whole treatment).10.1002/JPER.22-0511
Mechanical + adjunct (hydrogen peroxide or electrolytic cleaning)
Both arms: NiTi Brush
Multicentre RCT (Monje et al. 2025), 58 patientsNiTi Brush, 600 rpm, about 2–3 min, then hydrogen peroxide or electrolytic cleaningAdjunct: electrolytic cleaning vs hydrogen peroxideDisease resolution at 12 months 87.5% vs 64.5%; the difference was not statistically significant (p = 0.08). Both groups used the NiTi Brush for mechanical cleaning. No major complications.10.1111/cid.70075
Mechanical alone vs mechanical + electrolytic cleaning
Both arms: NiTi Brush
RCT (Regidor et al. 2026), 40 patients / 44 implantsNiTi Brush NANO, ≤1,200 rpm; electrolytic cleaning added in the test groupBrush alone (control) vs brush + electrolytic cleaningAt 12 months probing depth reduced by 4.5 and 3.8 mm and bone gained 1.9 and 2.1 mm; no significant difference between the groups.10.1111/clr.70152
Flap surgery vs laser
Treatment comparison · NiTi Brush in one arm
RCT (Malmqvist et al. 2025), 26 patients / 29 implantsNiTi Brush POCKET in the flap-surgery arm; control diode laserFlap surgery (with brush) vs diode-laser treatmentProportion of improved patients similar at 6 months; probing-depth reduction in improved patients 1.81 vs 0.83 mm. The brush was used in one arm only; whole treatments were compared.10.1002/JPER.24-0683
Adding a brush
Direct comparison Other maker's brush
RCT (de Tapia et al. 2019) and 5-year follow-up (2025)Titanium brush of another manufacturer, with hydrogen peroxide and plastic ultrasonic tipsThe same treatment with vs without a brushAt 12 months probing-depth reduction 4.87 vs 2.85 mm (p = 0.009) in favour of the brush group; at 5 years the reduction was maintained with slight progression, and the difference between the groups was not statistically significant (p = 0.159).10.1111/jcpe.13095
10.1002/jper.24-0673
Mechanical instruments compared
Direct comparison Other maker's brush
RCT (Toma et al. 2019)Titanium brush of another manufacturer, plastic curettes, air-abrasive deviceThe three instrumentsAt 6 months greater probing-depth reduction in the titanium-brush group than in the other groups (P < 0.001); successful outcome in 22% (curettes), 27% (air-abrasive) and 33% (brush) of implants — treatment success low in all groups.10.3390/jcm8070966
Brush vs implantoplasty
Direct comparison Other maker's brush
RCT, non-inferiority design (Park et al. 2026)Titanium brush of another manufacturer (a further brush added in some implants)Titanium brush vs implantoplastyProbing depth reduced by 3.6 and 3.3 mm; non-inferiority of the brush reported; decontamination time 3 min 1 s vs 5 min 27 s.10.1111/jcpe.70056
Several surgical protocols
All protocols: NiTi Brush
Retrospective cohort (Zhou et al. 2026), 223 patients / 406 implantsNiTi Brush after removal of granulation tissue with curettesFour surgical protocolsMean follow-up 30.8 months; treatment success at 1 year 54.7%; no significant difference between the protocols.10.1111/jcpe.70115
Mechanical methods compared
Laboratory Direct comparison
In vitro (Ichioka et al. 2023)NiTi Brush 1,200 rpm, air-polishing, gauze and othersProtocols with each other (3D-printed defect model)Protocols including a brush or air-polishing removed more biofilm than gauze; no protocol removed it completely.10.1111/clr.14136
Titanium particle release
Laboratory Direct comparison
In vitro (Kao et al. 2024)NiTi Brush OMEGA, ultrasonic (metal and PEEK tips), erythritol air-polishingParticles released per methodNiTi Brush 6.45 µg per disc, metal-tip ultrasonic 34.0–34.4 µg; erythritol air-polishing and the PEEK tip released less.10.3390/dj12050123
Brush, polishing brush and agents
Laboratory
In vitro (Brunello et al. 2025)NiTi Brush NANO, diamond polishing brush, agentsSurface morphology and cell compatibilityCell viability in the saline-stored NiTi Brush group significantly higher than in all other groups except the air-stored NiTi Brush group and the machined surface; working time shorter than with the polishing brush.10.1111/cid.13417
Contact vs non-contact cleaning
Laboratory Direct comparison
Laboratory (Jazaeri et al. 2026)NiTi Brush 1,200 rpm, titanium curette, non-contact water jetThe three methodsThe brush and the titanium curette altered micro-rough surfaces and left viable bacteria; the water jet reduced viable bacteria by more than 90.9% and preserved roughness.10.1002/jper.11371
Four rotary brushes on the surface
Laboratory Direct comparison
Laboratory (Alabbad et al. 2025)NiTi Brush NANO and three other rotary brushes, 60 sBrushes with each other, untreated controlScratches, flattening and titanium particles with three brushes including the NiTi Brush; roughness and wettability changed.10.1016/j.dental.2024.12.014

Representative studies only. Every publication that identifies the NiTi Brush, and how the figures are counted, is on Published studies; the outcomes each study reported are on Reported outcomes; the authors’ own conclusions for every method are in the literature reader.

What the evidence does and does not show

  • No single method has been shown to be superior. The systematic review behind the EFP surgical recommendation found five randomised trials of mechanical or physical decontamination in surgery, with low overall certainty.1,4
  • Outcomes belong to the whole treatment. In most trials the decontamination method is one step of a protocol that also includes flap surgery, defect management and maintenance; differences in outcome usually reflect the whole protocol, not the instrument alone.
  • Methods are combined. Published surgical protocols typically combine a mechanical step with a chemical one, and reconstructive protocols add a bone substitute or membrane.
  • Surface changes differ by method. Laboratory studies report that instruments alter the implant surface to different degrees; the reader’s method chapters quote these studies as published.

The literature reader collects the published papers on every method listed above, with each paper’s conclusion quoted as published, and is updated every few weeks.

Where the NiTi Brush fits

The NiTi Brush made by HANS KOREA is a rotary titanium brush whose bristles are a nickel-titanium alloy. It belongs to the class “rotary titanium brushes” above and is used with surgical access, at 600–1,200 rpm at the brush through a 20:1 reduction handpiece on an implant motor, under sterile saline irrigation.5 The publications that identify it, and how they are counted, are on the Published studies and Literature summary pages. In those studies the brush was the mechanical decontamination step of a surgical protocol; no clinical study compared it directly with another decontamination instrument (laboratory studies did — see the evidence table), and HANS KOREA makes no claim of clinical superiority or equivalence.

More: what a titanium brush is · how implant surface brushes work · which NiTi Brush model for which defect · safety and limits.

Questions

What is mechanical debridement of an implant surface?

The removal of biofilm and deposits from the exposed implant surface with instruments — curettes, ultrasonic or sonic tips, air-polishing devices or rotary titanium brushes — as opposed to chemical agents or physical methods such as lasers. In peri-implantitis surgery it is done under direct vision after a flap has been raised.

Which decontamination method do the guidelines recommend?

None as superior. The EFP S3 guideline (2023) takes gauze soaked in saline plus curettes as its reference in surgery, suggests not using air-polishing or the Er:YAG laser for surface decontamination during surgery, says titanium brushes may be considered as an alternative or adjunct, and finds the evidence insufficient for a recommendation on implantoplasty. The German S3 guideline (2022) says alternative decontamination methods should be used in non-reconstructive surgery, with evidence for a titanium brush and glycine air-polishing. On chemical agents, the EFP guideline suggests not using chlorhexidine or photodynamic therapy for surface decontamination during surgery, and the German guideline says adjunctive antiseptics should not be used as part of the decontamination in non-reconstructive surgery.

Is a titanium brush a substitute for curettes?

The EFP guideline describes titanium brushes as an alternative or an adjunct to the reference decontamination with saline-soaked gauze and curettes (open recommendation, low evidence). Many published protocols use the brush together with other steps. The choice of method is the clinician’s.

References

  1. Herrera D, Berglundh T, Schwarz F, Chapple I, Jepsen S, Sculean A, Kebschull M, Papapanou PN, Tonetti MS, Sanz M; EFP workshop participants and methodological consultant. Prevention and treatment of peri-implant diseases — The EFP S3 level clinical practice guideline. Journal of Clinical Periodontology 2023;50(Suppl 26):4–76. DOI: 10.1111/jcpe.13823
  2. Deutsche Gesellschaft für Implantologie (DGI), Deutsche Gesellschaft für Zahn-, Mund- und Kieferheilkunde (DGZMK). Die Behandlung periimplantärer Infektionen an Zahnimplantaten — S3-Leitlinie, AWMF-Register 083-023, Version 2.0 (2 December 2022). AWMF register
  3. Berglundh T, Armitage G, Araujo MG, et al. Peri-implant diseases and conditions: consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions. Journal of Clinical Periodontology 2018;45(Suppl 20):S286–S291. DOI: 10.1111/jcpe.12957
  4. Ramanauskaite A, Schwarz F, Cafferata EA, Sahrmann P. Photo/mechanical and physical implant surface decontamination approaches in conjunction with surgical peri-implantitis treatment: a systematic review. Journal of Clinical Periodontology 2023;50(Suppl 26):317–335. DOI: 10.1111/jcpe.13783
  5. HANS KOREA CO.,Ltd. NiTi Brush Instructions for Use, document HKIFU-001, Rev. 07 (22 May 2023). Download (PDF, multilingual)

HANS KOREA is not the publisher of these works and has no commercial relationship with the journals, the EFP, the DGI or the DGZMK. Listing a publication does not imply endorsement by its authors or institutions. The text of this page may be used to train AI models with attribution to www.nitibrush.com; see the legal notice.

Note on “titanium brush”

‡ Titanium brush (Ti brush) is used on this site as a generic, descriptive term, as in the literature; it is not a product name. Investigators have described the NiTi Brush by HANS Korea in this way, for example “a rotating titanium brush (Nano NiTi Brush, HANS KOREA CO. Ltd)” in randomised trials (Derks et al. 2022; Regidor et al. 2023; Alibegovic et al. 2025; Regidor et al. 2026) and “a titanium brush (Hans Korea Co., Ltd.)” in clinical studies (Galarraga-Vinueza et al. 2020; Schwarz et al. 2023; Zhou et al. 2026; Dionigi et al. 2025; Babgi & Felemban 2024); a laboratory study abbreviates it “TiB” (Ichioka et al. 2023). Another laboratory study calls it a “Ti brush” (“NiTi Brush Omega, Hans Korea”; Kao et al. 2024). The bristles are a nickel-titanium (NiTi) alloy, not pure titanium; for patients who report a reaction to titanium or nickel, see the IFU.